Runs Digest

The following is a digest of the T9 runs. Get the full list or the selected run from the pipeline or elogbook. (WORK IN PROGRESS).

In red color : Not to be used, due to mistake, or for specialist. 

In blue color the runs numbers good for the statistics 

Last run taken into account : 1334

Summary of the Digest

 CU Position 1 (0°)

   Electron Full Brem   Tagged
 0.5 GeV
  20k (1204)
  275 k (1244-1258)
 1 GeV
 20k (1205)   125 k (1208-1210)
 1.5 GeV
    50 k(1332, 1334)
 2.5 GeV
60k (1201-1202 )
 2,5 M (1181-1190) 100 k (1177-1180)

 CU Position 2 (30°)

  Electron Full Brem   Tagged
 0.5       125k(1295, 1297-1300)
+ 60k(1303-1305)
 1. GeV
 100k (1220)
  105 k (1214-1217)
+50 k(1310-1311)
1.5 GeV       
 2.5    2.5 M(1223-1231)  105k (1233-1237)

 CU position 3 (50°)

   Electron Full Brem   Tagged
0.5   
 
   75 k+85 k =160 k
(1287_1289)(1291-1294)
 1.  
 
   100k (1276-1279)
50k(1308-1309)
 1.5
     
 2.5  
 
200k (1262-1271)  100k (1272-1275)

 CU Position 4 (145°) x=749, y=0, z= -175 -215°

   Electron Full Brem   Tagged
 0.5     35k(1329-1330)
 1.     50K (1319-1320)
50k (1327-1328
 2.5    2Mo (1312-1318)  

Electron runs

 Runs  Date particle
energy
CU configuration
CU position
AD
notes
 1201  11/8 e-   2.5  BT1  0°    av rate 34 HZ, CU instant rate 500 hz. S0 counts 21k per spîll
 1202    e-  2.5  BT1      3 phasing errors for 30k events. run stopped due to beam off
 1204  11/8  e-  0.5  
 
     20k
1205

e-
1.



S0:14 k count per spill
HW trg: 700 count per spill
DAQ trg: 300 count per spill
beam divergence: 6.3mrad on XZ, 5mrad on YZ
 1218-1219 11/08 
21h43 
 e-  1.    30° (x,y,z)=(200,13.2,-48) no   Accidentally started runs in synchronized Run Control. These were automatically aborted by the program.
 1220  22h06  e- 1.  BT1  idem   No  100K, S0 new trigger, S4 off
 1221  22h20  e-  1. BT1  Idem   YES  Refrence point before changing momentum
 1239  12/08
11:02 
 e-  .5    idem    500 MeV electron with magnet off to check dump position with electron beam.9K
 1240
   .5    idem    500 MeV electron with magnet off to check dump position with electron beam.
 1259  12/8
21:17
 e'  1.        1. Gev with magnet off
 1260  12/8
21:34 
  2.5        magnet off, 2.5 GeV electrons to check beam and CU positions
 1261  12/8
21:53 
   2.5        Magnet Off
XY table has been moved few centimeters

 

Tagged photon runs

 Runs  Date particle
energy
IMagnet  CU configuration
CU position
AD
notes
1176    tagged g
3GeV
 600 BT1
0 degrees, center of tower 2
yes  events in twr3?
 1177-1180   tagged g
e beam at 2.5GeV
 600 BT1
0 degrees, center of tower 2
yes
magnet ON, 25K evts per run 
 1206    tagged gamma
e-  1GeV  240     idem    S0:14k count per spill
HW trg: 100 count per spill
DAQ trg: 100 count per spill
Lot of 0 cluster in chamber 2 and 3, we need to tune the bending magnet and/or improve the veto to select a small spot.
 1207  11/8  tagged gamma     240     idem    idem with S1 in the trigger
this run is to see if the zero that we see is a trigger or magner problem: it seem that it's a trigger issue: now the fraction of zero seems similar to what we saw in the other physics runs. 
 1208 11/8 11h34
 tagged gamma  e-
1 GeV
 240     idem yes
  New trigger S3 in veto  - 25k
 1209  11/8
13h29 
 tagged gamma  e-
1GeV
 240    idem  yes and no
around 30k : SYNCHRONIZATION LOST !
stopping the run, first 30000 events should well sycnhronized...  
1210   11/8
16h34 
 tagged gamma  e-
1GeV 
 240    idem  yes  Synchronized Tagged gamma with 1GeV electron beam, going for 70k to get all the statistic : 100Kevents
(+25 k from 1208 this cpunt is not taken into account in the eloogbook)
 1211-1213  11/8   Tagged  1 GeV 180-240
  BT1  30° (x,y,z)=(200,13.2,-48)  
 various Magnet currents for setting up
 1214  11/8  Tagged  e 1 GeV
 240  BT1  30° (x,y,z)=(200,13.2,-48)     25k
 1215    Tagged  e 1GeV  240  BT1  idem    25k
 1216    Tagged  e 1 GeV
 240  BT1  idem    25K
 1217  11/8
21h22 
 Tagged  e- 1 GeV
 240  BT1  idem    30k
No spill during 4 minute
 1232  12/08  Tagged  e- 2.5 GeV
 600   BT1  30° (x,y,z)=(200,13.2,-48)    Not to be used , S4 not in the Trigger
 1233-1237 12/8 
7:21-9:36 
 Tagged  e- 2.5
 600  BT1  idem    
 1242  12/8
11:45 
   e- 0.5 GeV
 100     0 degrees, center of tower 2    GeometryVerifiation
1243    Tagged    e- 0.5 GeV idem      0 degrees, center of tower 2  yes  rate too low
 1244-1245 12/08
12:44 13:35 
 Tagged   e- 0.5 GeV  120     0 degrees, center of tower 2  yes  
 1246-1250 13:57   Tagged   e- 0.5 GeV 120     0 degrees, center of tower 2  yes and no
 Sync lost after 6476 evts
1253-1258     tagged e 0.5gev
 120     0 degrees, center of tower 2  yes  
1272
13/8
3:52 
Tagged
e 2.5 GeV
600

48°

 Table moved to 48 deg, in order to allow full calrecon  
 1273-1275  13/8
Tagged  e 2.5 GeV  600     48°    table@(-50,12.3,-48, 48deg)  
 1276-1279  13/8  Tagged  e 1GeV
 240    48   idem   
 1287-1289  13/8
11:07-12:41 
 Tagged  e 0.5
 120    48°
 Table(-50,12.3,-48, 48deg)
Trigger S0- S2- S4 -veto Sh - veto S3 - C1 - C2
 
 1290                Junk  
 1291-1294  13/8
11:07-12:41 
 Tagged  e 0.5
 120    48°
  85 k  
 1295,  13/8
16:22
 Tagged  e 0.5
120     30 °
  25k   
 1297-1300  13/8
18:11-19:10
 Tagged  ie 0.5  idem    30°     100 k
 
 1303  13/8  Tagged  e 0.5
 idem    30°    25k  
 1304  13/8  Tagged e 0.5   120    30°    12 k . Rinsing number of events out of Sync
 
 1305 13/8   Tagged  e 0.5
 120   30°
   25k  
 1308-1309  14/8
3:13 
 Tagged e 1GeV   240   48°    50K
in previous run 1307, Moved tile 4 (attached to GAFE 7) from on top of tracker tower to the +X side of the CU, at the base of the tower
This position corresponds to the tile 330 in the LAT.
 
 1310-1311 14/8  Tagged  e 1 GeV  240    30°    50k   
 1319-1320    Tagged  1 GeV
240    Pos 4, 145°   Yes 
 50k. 
 
1321 14/8          idem   Yes and no
 sync lost after ??
 
 1327-1328 14/8  Tagged   1 GeV
240     idem Yes  50k   
 1329-1330  14/8
12:16 
Tagged   0.5 GeV
120     Pos 4, 145°
 yes 35k   
 1331      1.5 GeV
360    Pos 1, 0°
 No  test beam position, no sync
 
 1332 14:27  Tagged  1.5 GeV  360    Pos 1, 0°  Yes
 25k  
 1333    Tagged 1.5  360    same   yes and no
 no sync after <3600 evts
 
 1334-1335 14/8
16:12
 Tagged  1.5 GeV
360    Same   yes 50 k  
 1336, 1337
           same   sync lost afyer 1733 and 1280 evts   
 1338            same   25k   
 1339-1342            Pos 2
   100k  
 1343-1346                  

Full-brem Photon runs

 Runs  Date particle
energy
 I(Magnet) CU configuration
CU position
AD
notes
1181-1190  8/8
9/8 
e+g
2.5
 600 BT1
Pos 1 : 0 degrees, center of tower 2
no overall 2M evts  - (false report of 300 A for 1185 in elogBook)
1222
 11/8

  2.5 GeV
 0
Pos 2 : 30° (x,y,z)=(200,13.2,-48 yes
check beam position
 1223-1231 12/8  e+g  2.5 GeV   600  BT1 idem   no  rate stable after adjustement in run 1223
2.5 Mevents 
 1262-1271  13/8
<3:02 
 e+g  2.5  600     pos 3 : 50°    
 1312-1318  14/8    2.5 600    Pos 4 : 145 °      2 Mevents

Tagger calibration runs

 Runs Date  particle
energy (GeV)
CU configuration
CU position
AD
notes
 1148-1164; 1166-1169  8/8 e beam
various
BT1
0 degrees, various positions to align with tagger
yes
many different configurations, for expert only
 1170   e beam
1.5
BT1
0 degrees, center of tower 2  yes  dump in place, slits at 51.5
 1171   e beam
1.25 
BT1
0 degrees, center of tower 2  yes magnet current 600A, direct tagger calibration
 1172   e beam
1.125
BT1 0 degrees, center of tower 2  yes magnet current 600A, direct tagger calibration-NOT Synchronised
 1173   e beam
 1.375 BT1
0 degrees, center of tower 2
 yes magnet current 600A, direct tagger calibration, slits reduce to 51
 1174  8/8 e beam
1.125 
 BT1 0 degrees, center of tower 2
yes
magnet current 600A, direct tagger calibration
1282
13/8
e

BT1
48°,

Test - - Tagger calibration
Trigger is standard : S0.S2.S4.C1.C2.veto(S3+Sh)
Magnet Current set to 300A to shoot the beam directly in the Tagger.
300A is not enough trying to go up to 320A and then to 350A and then to 400A
400A is OK, Stopping run and start the Good On
 1283  13/8  e  1 eEV
 BT1      400 A
 1284  13/8  e  0.75  BT1      330 A Synchronization Lost from the beginning...
 1285  13/8  e 0.75 GeV BT1       Magnet Current = 330A
All 4 Si tagger chamber are fully populated.
 1286 13/8  .5         Magnet current @ 120 A
Pressure of C1 = 0.9 bars, C2=0.97 bars
Beam shut down after 18553 events


 

 CAL calibration runs

 Runs particle
energy
CU configuration
CU position
AD
notes
             
             
             
             

CAL pedestal runs

 Runs Date  particle
energy
CU configuration
CU position
AD
notes








 1165  8/8      BT22(S3 only)
     pedestal
 1191  9/8 8h51
-
-
 BT22 (S3 only)
-
-
monitor pedestal shift
 1198  10/8      BT22      pedestal T ~22.3C
 1201  11/8      BT22      Pedestal T~ ?
 1281  13/8            Pedestal- Trigger is now only S0 and the beam is off...have only muons deviated by the magnet.
We used to trigger on S3 when it was out of the beam, but now it's on the beam line, so we had to find a new strategy !
Average Rate 26Hz
 1348  14/8
21:26 
    BT 22       5k

ACD Calibration runs

Runs Date  particle
energy
CU configuration
CU position
AD
notes
 1192, 1193
 9/8  cosmic          
 1194-1197 10/8   cosmic    BT3      

 

Sil Calibration runs

Runs Date  particle
energy
CU configuration
CU position
AD
notes
 4318  11/8            

Cosmic run

Runs Date  particle
energy
CU configuration
CU position
AD
notes
 1307              muons to check light leak

 

Not to be used 

Runs Date  particle
energy
 IMagnet CU configuration
CU position
AD
notes 
 1200  11/8  e-  2.5  0        ? Control Room ajustingSpill structure during run
 1203   11/8  e- 0.5    0      
 S0 = 13k/spill, wrong focus setting









 1238  12/8              Software test
 1241  12/8              Software test
 1280  13/8              Test run for CU CAL pedestals - JUNK
 1296                Junk
 1301-1302                Junk
1322-1326                junk
 1347                first run to prepare positron setting of tonight
Turn Off Magnet and Move table to position (-50, 0, 550, -45degree)
 1349-1354                to check a lead brick position that should simulate the MMS.
  • we start with 240 A that is too much: the beam miss the lead and enter the CU side.
  • second setting is 200A, it seems better. let's scan more current to be sure.
  • magnet 160 A, beam is bent too much... etc..

 